QCM Field Guide
A field guide for PhD students

The quartz crystal microbalance

A 334 µm slice of quartz, shaking sideways five million times a second, can weigh a layer of molecules thinner than a nanometre. This guide covers the physics, the models, and the judgement calls you need to turn a frequency shift into a number you can defend.

AT-cut θ = 35°15′ f0 5 MHz tq ≈ 334 µm C 17.7 ng cm⁻² Hz⁻¹ Zq 8.8×10⁶ kg m⁻² s⁻¹
Thickness-shear mode
f₁ = 5.000 MHz · λ/2 = tq Displacement exaggerated ~10⁵×
01 · PHYSICS

The resonator

Quartz is piezoelectric. An AC voltage across gold electrodes on both faces of an AT-cut disc drives a thickness-shear vibration: the two faces slide in opposite directions, parallel to the surface. The disc resonates when its thickness holds an odd number of half shear wavelengths.

The AT cut (rotated Y-cut at 35°15′) is chosen because its frequency–temperature curve has a turning point near room temperature, so thermal drift of the bare crystal is small.

Only odd overtones (n = 1, 3, 5 …) can be driven. For even n, the strain integrated over the thickness vanishes, so the electrodes see no net charge. In practice one tracks n = 3 to 13, because the fundamental is most sensitive to mounting.

Resonance condition \[ f_n \approx n\,f_0 = \frac{n\,v_q}{2\,t_q} = \frac{n}{2t_q}\sqrt{\frac{\mu_q}{\rho_q}}, \qquad n = 1,3,5,\dots \] vq ≈ 3340 m/s, so a 5 MHz crystal is ≈ 334 µm thick.
Standing shear wave \[ u(y,t) = u_0 \cos\!\left(\frac{n\pi y}{t_q}\right) e^{i\omega t} \] Antinodes sit on both faces. That is why anything attached to the surface is dragged along and changes f.
Acoustic impedance of AT-cut quartz \[ Z_q = \sqrt{\rho_q\,\mu_q} = 8.8\times10^{6}\ \mathrm{kg\,m^{-2}\,s^{-1}} \]

Where the crystal is sensitive

Radial profile

Energy trapping confines the vibration under the electrode. Mass sensitivity follows the squared amplitude, which is roughly Gaussian and peaks at the centre. Higher overtones are more tightly confined. Sauerbrey’s equation assumes a uniform film, so patchy deposits near the edge are under-weighted and deposits at the centre are over-weighted. The curves are schematic.

n = 1n = 3n = 5electrode edge
02 · MEASUREMENT

Reading a resonance

Electrically, the crystal near resonance behaves like a Butterworth–Van Dyke circuit: a motional branch (R1, L1, C1) in parallel with the static electrode capacitance C0. Mass loading adds inductance. A viscous load adds both inductance and resistance.

Every measurement technique extracts two numbers: the resonance frequency f and its width. Impedance analysers fit the conductance peak and report the half-bandwidth at half maximum Γ. Ring-down instruments (QCM-D) switch the drive off and fit the decay time τ, reported as the dissipation D. They measure the same physics.

It is convenient to combine both into one complex frequency shift, Δf̃ = Δf + iΔΓ. All the load models below predict Δf̃.

BVD equivalent circuit R₁L₁C₁C₀
Admittance and resonance \[ Y(\omega) = i\omega C_0 + \frac{1}{R_1 + i\omega L_1 + (i\omega C_1)^{-1}},\qquad f_s = \frac{1}{2\pi\sqrt{L_1C_1}} \]
Conductance near resonance \[ G(f) \approx G_{\max}\,\frac{\Gamma^2}{(f-f_r)^2+\Gamma^2} \] Γ is the half-width at half-maximum.
Dissipation, Q and ring-down \[ D = \frac{1}{Q} = \frac{E_{\text{lost per cycle}}}{2\pi E_{\text{stored}}} = \frac{2\Gamma}{f_r} = \frac{1}{\pi f_r \tau} \] From A(t) = A₀ e−t/τ sin(2πfrt + φ).

Load the crystal and watch the peak

Interactive
bare crystal in airloaded
Each peak normalised to its own maximum. Real peak height scales as 1/Γ.
bare envelopeloaded decay (carrier not to scale)
Time domain view of the same resonance. A wider peak means a faster decay.
Δfn
Δfn/n
ΔΓn
D (loaded)
Q (loaded)
τ (loaded)
03 · RIGID FILMS

The Sauerbrey regime

Sauerbrey (1959) treated a thin rigid film as extra quartz. The film moves with the surface and adds to the vibrating mass without storing or dissipating energy. The relative frequency drop then equals the relative mass gain:

\[ \frac{\Delta f}{f_0} = -\frac{\Delta t_q}{t_q} = -\frac{\Delta m_A}{\rho_q t_q} \]

Substituting tq = vq/2f0 and generalising to overtones gives the working equation. It is linear, needs no film properties, and is the reason the QCM is called a microbalance.

Sauerbrey equation \[ \Delta f_n = -\frac{2\,n\,f_0^{2}}{\sqrt{\rho_q\mu_q}}\,\Delta m_A = -\frac{n}{C}\,\Delta m_A \] ΔmA is mass per area. Report Δfn/n so overtones are comparable.
Mass sensitivity constant \[ C = \frac{Z_q}{2 f_0^{2}} = 17.7\ \mathrm{ng\,cm^{-2}\,Hz^{-1}} \ \ (f_0 = 5\ \mathrm{MHz}) \] C scales as 1/f₀²: 4.42 ng cm⁻² Hz⁻¹ at 10 MHz.
Complex form (small-load) \[ Z_L = i\omega m_A \;\Rightarrow\; \frac{\Delta\tilde f}{f_0} = \frac{i Z_L}{\pi Z_q} = -\frac{\omega\, m_A}{\pi Z_q}, \quad \Delta\Gamma = 0 \]
ThinFilm mass is small compared with the crystal: |Δf|/f₀ below about 2 %.
RigidNo shear deformation across the film: ΔΓn ≪ |Δfn| and Δfn/n equal for all n.
UniformEvenly spread over the electrode, since sensitivity is radially weighted (section 01).
CoupledNo slip at the interface. In liquid, the mass includes water that moves with the film.

Sauerbrey calculator

Tool
C at this f₀
ΔmA
Film thickness
Mass on electrode
|Δf|/fn
Beyond ~2 %: Lu–Lewis Z-match (thick rigid films in vacuum) \[ m_A = \frac{N_q\,\rho_q}{\pi f_c\,R_Z}\,\arctan\!\left[R_Z \tan\!\left(\pi\,\frac{f_q-f_c}{f_q}\right)\right],\qquad R_Z = \frac{Z_q}{Z_f},\quad N_q = f_q t_q = 1670\ \mathrm{m\,s^{-1}} \] fq: unloaded frequency, fc: loaded. Used in thin-film deposition monitors. Reduces to Sauerbrey for small loads or when RZ = 1.
04 · NEWTONIAN LIQUIDS

Crystals in liquid

Immerse one face in a liquid and the shear wave leaks into it, decaying over the penetration depth δ. The liquid layer within δ is dragged along (a frequency drop) and sheared (energy loss, so Γ and D increase).

For a semi-infinite Newtonian liquid, Kanazawa and Gordon (1985) found that Δf and ΔΓ are equal in magnitude and grow with √n. Changing buffer, salt concentration or temperature therefore shifts f even when nothing adsorbs. Always compare against a KG estimate before you interpret a baseline step as binding.

δ is also the sensing range. At 5 MHz in water it is about 250 nm, which is why the QCM probes the interface rather than the bulk. Higher overtones probe a thinner slice.

Kanazawa–Gordon \[ \Delta f_n = -\Delta\Gamma_n = -\sqrt{n}\;f_0^{3/2}\sqrt{\frac{\rho_L\,\eta_L}{\pi\,\rho_q\,\mu_q}} \] Water at 5 MHz, n = 1: Δf ≈ −714 Hz.
Penetration depth and velocity field \[ \delta = \sqrt{\frac{2\eta_L}{\rho_L\,\omega}},\qquad v(z,t) = v_0\,e^{-z/\delta}\cos\!\left(\omega t - \frac{z}{\delta}\right) \]
Load impedance of the liquid \[ \tilde Z_L = \sqrt{i\omega\rho_L\eta_L} = (1+i)\sqrt{\frac{\omega\rho_L\eta_L}{2}} \] Equal real and imaginary parts give |Δf| = ΔΓ.

The shear wave in the liquid

Interactive
velocity v(z, t)envelope ±e−z/δz = δ
δ
Δfn (KG)
Δfn/n
ΔDn
05 · VISCOELASTICITY

Soft films

Polymer brushes, protein layers, lipid vesicles and hydrogels are not rigid. The film deforms in shear, so part of its mass lags behind the crystal and energy is lost. Three signatures tell you the Sauerbrey equation has stopped working: ΔΓ (or ΔD) is no longer small relative to −Δf, the normalised shifts Δfn/n spread between overtones, and the apparent mass depends on n.

The general tool is the small-load approximation. When the load is small compared with the crystal, the complex frequency shift is proportional to the load impedance Z̃L at the surface. A film of thickness d on top of a liquid is handled like a transmission line. Model the film as Kelvin–Voigt (spring and dashpot in parallel), or give G̃ directly.

In liquid, a soft film shows the missing-mass effect: −Δf/n under-reports the mass. In air, a soft film of finite thickness shows the opposite, called film resonance enhancement. The film section of the widget below computes both from the full layer formula.

Small-load approximation \[ \frac{\Delta\tilde f_n}{f_0} = \frac{\Delta f_n + i\,\Delta\Gamma_n}{f_0} = \frac{i\,\tilde Z_L(\omega_n)}{\pi Z_q} \] f₀ here is the fundamental. ΔDn = 2ΔΓn/(n f₀).
Film on a liquid (transmission line) \[ \tilde Z_L = \tilde Z_f\,\frac{\tilde Z_{liq} + i\tilde Z_f\tan(\tilde k_f d)}{\tilde Z_f + i\tilde Z_{liq}\tan(\tilde k_f d)},\quad \tilde Z_f = \sqrt{\rho_f \tilde G_f},\quad \tilde k_f = \frac{\omega\rho_f}{\tilde Z_f} \]
Kelvin–Voigt film \[ \tilde G_f = G' + i\,G'' = \mu_f + i\,\omega\,\eta_f \]
Thin-film limits (relative to the bare medium) \[ \text{liquid: }\ \frac{\Delta\tilde f}{f_0} \approx -\frac{\omega m_f}{\pi Z_q}\left(1 - \frac{\tilde Z_{liq}^{2}}{\tilde Z_f^{2}}\right) \qquad \text{air: }\ \frac{\Delta\tilde f}{f_0} \approx -\frac{\omega m_f}{\pi Z_q}\left(1 + \frac{\omega^2\rho_f d^2}{3\,\tilde G_f}\right) \]

Overtone fingerprint of a film

Interactive · full layer model
Presets:
−Δfn/nSauerbrey (true mass)
ΔDn (×10⁻⁶)
True areal mass
Sauerbrey mass from n = 3
Apparent / true
ΔD₃ / (−Δf₃/3)
Rigid if ΔDn/(−Δfn/n) ≪ 2/f₀ = 4×10⁻⁷ Hz⁻¹ (equivalently ΔΓn ≪ −Δfn). The reference state is the bare crystal in the same medium. The film is modelled as one homogeneous Voigt layer.
Fitting in practice. Fit Δfn and ΔΓn (or ΔDn) across at least three overtones at once. A Voigt layer has four unknowns (d, ρf, μf, ηf), and ρf and d are strongly correlated, so fix ρf. Thin soft films are often underdetermined: if the fit is flat along some direction, report a bound, not a value. Voinova et al. (1999) give the classic Voigt treatment. Johannsmann (2015) gives the power-law generalisation.
06 · DECISIONS

Choosing a model

Answer a few questions about your data. The guide suggests the simplest model that is still defensible.

Electrochemical QCM (EQCM) \[ \Delta m_A = -\frac{C}{n}\,\Delta f_n, \qquad \mathrm{mpe} = \frac{F\,\Delta m_A\,A}{Q} = \frac{M}{z} \] Plot Δm against charge Q. The slope (mass per electron, g/mol) identifies the species inserted or deposited, e.g. 107.9 g/mol for Ag⁺ + e⁻.
Hydration from a combined experiment \[ h = \frac{m_{\text{QCM}} - m_{\text{opt}}}{m_{\text{QCM}}} \] The QCM weighs coupled solvent. SPR, ellipsometry or reflectometry see dry mass. Their ratio gives the solvent fraction of the layer.
07 · BENCH

Lab practice

EffectTypical size (5 MHz)What to do
Temperature in water (through η(T))≈ 7–8 Hz/K at n = 1Thermostat the cell to ±0.02 K. Let it equilibrate 30–60 min before the baseline.
Bulk solution change (buffer, salt, glycerol)KG: ∝ √(ρη)Run a reference crystal or compute KG shifts. Rinse back to the original buffer to separate bound mass from bulk effects.
Noise floor~0.1 Hz ≈ 2 ng/cm²Average, use n ≥ 3, and report the noise with every LoD.
Mounting stress, O-ring pressuretens of Hz, mostly n = 1Use a consistent mounting torque. Trust overtones over the fundamental.
Air bubbles on the sensorsudden jumps in f and DDegas buffers, prime the flow cell, inspect before injecting.
Hydrostatic pressure or flow rateHz-level stepsKeep the flow rate constant through the whole run. Do not stop the pump just to inject.
Surface roughness and porositytrapped liquid adds massCharacterise with AFM. Compare with a smooth reference sensor.
Non-uniform coverageradial weighting errorCoat the whole active area. Check uniformity optically.
Unit conversions (f₀ = 5 MHz, ρ = 1 g/cm³) \[ 1\ \mathrm{Hz} \;\leftrightarrow\; 17.7\ \mathrm{ng\,cm^{-2}} \;=\; 0.177\ \mathrm{mg\,m^{-2}} \;\leftrightarrow\; 0.177\ \mathrm{nm} \] 1 ng/cm² = 0.01 mg/m² = 0.01 nm at unit density.
Reporting checklist. Crystal cut and f₀, overtones used, electrode material and area, temperature, flow rate, reference state, whether shifts are Δfn or Δfn/n, how D was measured (ring-down or impedance), the model used, and the fixed parameters in any fit.
08 · CHECK YOURSELF

Self-test

Six questions. Pick an answer to see the reasoning.

0 / 0
09 · REFERENCE

Constants & reading

QuantitySymbolValue
Density of quartzρq2648 kg m⁻³
Shear modulus, AT-cutμq2.947×10¹⁰ Pa
Shear wave speedvq≈ 3336 m s⁻¹
Acoustic impedanceZq8.8×10⁶ kg m⁻² s⁻¹
Frequency constantNq = f₀tq1670 m s⁻¹ (1.67 MHz·mm)
Mass sensitivity, 5 MHzC17.7 ng cm⁻² Hz⁻¹
Mass sensitivity, 10 MHzC4.42 ng cm⁻² Hz⁻¹
KG shift, water, 5 MHz, n = 1Δf≈ −714 Hz
Penetration depth, water, 5 MHzδ≈ 252 nm (÷√n)
  1. G. Sauerbrey, “Verwendung von Schwingquarzen zur Wägung dünner Schichten und zur Mikrowägung,” Z. Phys. 155, 206 (1959).
  2. K. K. Kanazawa, J. G. Gordon II, “The oscillation frequency of a quartz resonator in contact with a liquid,” Anal. Chim. Acta 175, 99 (1985).
  3. C.-S. Lu, O. Lewis, “Investigation of film-thickness determination by oscillating quartz resonators with large mass load,” J. Appl. Phys. 43, 4385 (1972).
  4. S. J. Martin, V. E. Granstaff, G. C. Frye, “Characterization of a quartz crystal microbalance with simultaneous mass and liquid loading,” Anal. Chem. 63, 2272 (1991).
  5. M. Rodahl, F. Höök, A. Krozer, P. Brzezinski, B. Kasemo, “Quartz crystal microbalance setup for frequency and Q-factor measurements in gaseous and liquid environments,” Rev. Sci. Instrum. 66, 3924 (1995).
  6. M. V. Voinova, M. Rodahl, M. Jonson, B. Kasemo, “Viscoelastic acoustic response of layered polymer films at fluid–solid interfaces,” Phys. Scr. 59, 391 (1999).
  7. I. Reviakine, D. Johannsmann, R. P. Richter, “Hearing what you cannot see and visualizing what you hear,” Anal. Chem. 83, 8838 (2011).
  8. D. Johannsmann, The Quartz Crystal Microbalance in Soft Matter Research, Springer (2015).